New Precision LCR Meter TH2827A

Feedback Ratings
  • 1-4 Sets
    $2,100.00
  • >=5 Sets
    $1,900.00
Quantity:
- + Sets available
0piecesselected, US $in total
View Details
Shipping fee:
Depends on the order quantity.
Lead Time:
2 day(s) after payment received
Contact Supplier
Report Suspicious Activity
Overview
Quick Details
Place of Origin:
Jiangsu, China (Mainland)
Brand Name:
Tonghui
Model Number:
TH2827A
Frequency:
300kHz
Basic accuracy:
0.01%
Test Speed:
13ms/s
Min.Resolution:
10mHz
Parameter:
L,C,X,B,R,G,D,Q,DCR

Packaging & Delivery

Selling Units:
Single item
Single package size: 
50X46X23 cm
Single gross weight:
7.0 kg
Package Type:
Carton Box
Picture Example:
package-img
Lead Time :
2

Introduction

TH2827A is an new impedance analyzer,LCR Meter, easy operation, accurate and stable reading.With test frequency from 20Hz to 300kHz, frequency resolution of 1mHz, wide test signal level range from 10mV to 2V, measurement speed up to 13ms/time, TH2827A can be used to inspect electronic components, not only for measurement parameters TH2819 has, but DCR measurement. The meter has ±5V (±50mA) and optional 1A internal DC bias source and external inductance bias sources to inspect inductive components.

With high speed, wide test range, and 6-digit resolution, TH2827A can be used in quality control on production line, incoming inspection and lab measurement.

Measurement function

Measurement parameters

|Z|, |Y|, C, L, X, B, R, G, D, Q, θ ,DCR

Basic accuracy

0.01%

Equivalents circuit

Serial, Parallel

Math function

Absolute value deviation, percentage deviation

Range mode

Auto, Hold, Manual

Trigger mode

Internal, Manual, External

Measurement time (≥1kHz)

Fast: 13ms, Medium: 67ms, Slow: 187ms

Average time

1—256

Delay time

0—999, with the step of 100μs

Correction

Open, Short, Load

Measurement terminal

5-terminal

Display mode

Direct, Δ, Δ%, V/I (voltage/current monitor)

Display

800×RGB×480 7 inch TFT LCD Display 

Test signal

Test signal frequency

20 Hz-300kHz

Frequency min. resolution

1mHz

Output impedance

30 Ω,  50Ω 100 Ω

Test signal level

Normal:       5 mV-2 V       Accuracy: 10%

Constant level: 10 mV-1 V      Accuracy: 5%

Level min. resolution

1mV

DC bias source

Internal

0V, 1.5V, 2V, accuracy: 1%

Optional

TH1902A 0-10V DC bias voltage, 0-100mA DC bias current

TH1902B 0-1A DC bias current

Measurement display range

|Z|, R, X

0.00001 Ω — 99.9999 MΩ

DCR

0.001 Ω — 99.9999 MΩ

|Y|, G, B

0.00001µS — 99.9999 S

C

0.00001 pF — 9.9999 F

L

0.00001 µH — 9999.99 H

D

0.00001 — 9.9999

Q

0.00001 — 9999.9

θ (DEG)

-179.999º — 179.999 º

θ (RAD)

-3.14159 — 3.14159

Δ%

-999.999% — 999.999%

Others

Comparator

10-bin sorting, bin counter

Storage

100 sets of LCRZ measurement settings saved

Interface

RS232C, HANDLER, LAN,GPIB (optional), USB HOST (optional)