Square wave characteristics: Overshoot: ≤5%(At 10Mv/DIV range) other distortions and other ranges:5% added to the above value
DC balance shift: Adjustable on panel
Linearity: ±0.1DIV of amplitude change when waveform of 2 DIV at graticule center is moved vertically.
VERTICAL MODES
CH1 single channel.
CH2 single channel
DUAL: CH1 and CH2 are displayed ALT or CHOP selectable at any sweep rate.
ADD: CH1+CH2 algebraic addition.
Chopping repetition frequency: Approx.250KHz
Input coupling: AC , GND , DC .
Maximum input voltage: 300V peak (AC: frequency 1KHz or lower) When set probe switch at1:1,the maximum effective readout is 40Vpp(14Vrms at sine wave),or set probe switch at 10:1,the maximum effective readout is 400Vpp(140Vrms at sine wave).
Common mode rejection ratio: 50:1 or better at 50KHz sinusoidal wave.(when sensitivities of CH1 and CH2 are set equally)
Isolation between channels: 1000:1 at 50 MHz (at 5mV/DIV range); 30:1 at 20MHz; 30:1 at 40MHz
CH1 signal output: at least 20mV/DIV into a 50 ohm termination. Bandwidth is 50Hz to at least 5MHz.
CH2 INV BAL: Balanced point variation: ≤1DIV (Reference at center graticule)
TRIGGERING
Triggering source: CH1,CH2,LINE,EXT(CH1and CH2 can be selected only when the vertical mode is DUAL or ADD. in ALT mode, if the TRIG.ALT switch is pushed in, it can be use for alternate triggering of two different source.