Main Specification | Wavelength range | 190-900nm |
Wavelength accuracy | ±0.25nm | |
Resolution | Two spectral lines of Mn at 279.5nm and 279.8nm can be separated with the spectral bandwidth of 0.2nm and valley peak energy ratio less than 30%. | |
Baseline stability | ≤0.004A/30min | |
Background correction | The D2 lamp background correction capability at 1A is better than 30 times. The S-H background correction capability at 1.8A is better than 30 times. (only for WFX-110A/120A) | |
Hollow Cathode Lamps | Lamp turret | 6-lamp turret (WFX-110A/120A) 4-lamp turret (WFX-130A) Auto-alignment, fully automated scan and peak-picking. |
Lamp current adjustment | Automatic adjustment and display. Wide pulse current: 0~25mA Narrow pulse current: 0~10mA | |
Lamp power supply mode | 400Hz square wave pulse; 100Hz narrow square wave pulse + 400Hz wide square wave pulse (WFX-110A/120A) | |
Optical System | Monochomator | Single beam, Czerny-Turner design grating monochromator |
Grating | 1800 l/mm | |
Focal length | 277mm | |
Blazed wavelength | 250nm | |
Spectral bandwidth | 0.1nm, 0.2nm, 0.4nm, 1.2mm, automatic change. | |
Flame Atomizer | Burner | 10cm single slot all-titanium burner |
Spray chamber | Corrosion resistant all-plastic spray chamber. | |
Nebulizer | High efficiency glass nebulizer with metal sleeve, sucking up rate: 6-7mL/min | |
Emission burner | Provided with WFX-110A/120A | |
Graphite furnace | Temperature range | Room temperature ~ 3000 |
Heating rate | 2000/s | |
Graphite tube dimensions | 28mm(L)´8mm(OD) | |
Characteristic mass | Cd≤0.8×10-12g, Cu≤5×10-12g, Mo≤1×10-11g | |
Precision | Cd≤3%, Cu≤3%, Mo≤4% | |
Detection and Data Processing System | Detector | R928 Photomultiplier with high sensitivity and wide spectral range. |
Software | Windows operating system. | |
Analytical method | Working curve auto-fitting; standard addition method; automatic sensitivity correction; automatic calculation of concentration and content. | |
Repeat times | Maximum 20 times of repeat measurement, automatic calculation of mean value, standard deviation and relative standard deviation. | |
Multi-task Functions | Sequential measurement of multi-elements in one sample. | |
Condition reading | With model function | |
Result printing | Measurement data and final analytical report printout, editing with Excel. | |
Standard RS-232 serial port communication | ||
Characteristic Concentration and Detection Limit | Normal Air-C2H2 flame | Cu: Characteristic concentration≤0.025mg/L, Detection limit≤0.006mg/L; |
Rich oxygen Air- C2H2 flame | Ba: Characteristic concentration≤0.22mg/L Al: Characteristic concentration≤0.4mg/L (for WFX-110A) | |
Function Expansion | Hydride vapor generator can be connected for hydride analysis. | |
Dimensions and Weights | 1020(L)´490(W)´540(H)mm(main unit), unpacked 80Kg | |
420(L)´420(W)´460(H)mm(graphite furnace), unpacked 50Kg | ||