Main Specification | Wavelength range | 190-900nm |
Wavelength accuracy | ±0.5nm | |
Resolution | Two spectral lines of Mn at 279.5nm and 279.8nm can be separated with the spectral bandwidth of 0.2nm and valley peak energy ratio less than 30%. | |
Baseline stability | ≤0.005A/30min | |
Background correction | The D2 lamp background correction capability at 1A is better than 30 times. | |
Light Source System | 2 lamps are powered simultaneously (one preheating). | |
Lamp current adjustment range:0~20Ma. | ||
Lamp power supply mode | Powered by 400Hz square pulse | |
Optical System | Monochomator | Single beam, Czerny-Turner design grating monochromator |
Grating | 1800 l/mm | |
Focal length | 277mm | |
Blazed wavelength | 250nm | |
Spectral bandwidth | 0.1nm, 0.2nm, 0.4nm, 1.2mm 4steps | |
Adjustment | Manual adjustment for wavelength and slit | |
Flame Atomizer | Burner | 10cm single slot all-titanium burner |
Spray chamber | Corrosion resistant all-plastic spray chamber. | |
Nebulizer | High efficiency glass nebulizer with metal sleeve, sucking up rate: 6-7mL/min | |
Position adjustment | Manual adjusting mechanism for vertical, horizontal positions and the rotation angle of the burner. | |
Gas line protection | Fuel gas leakage alarm. | |
Detection and Data Processing System | Detector | R928 Photomultiplier with high sensitivity and wide spectral range. |
Electronic and micro-computer system | Automatic adjustment of light source power. Light energy and negative high-voltage auto-balance. | |
Display mode | LED display of energy and measurement values, concentration direct reading. | |
Read mode | Transient, time average, peak height, peak area. Integral time is selectable in the range of 0.1-19.9s. | |
Scale expansion | 0.1~99 | |
Data processing mode | Automatic calculation of mean, standard deviation and relative standard deviation. Repeating number is in the range of 1-99 | |
Measurement mode | Automatic curve fitting with 3~7standars; Sensitivity auto-corrrection. | |
Result printing | Measurement data, working curve, signal profile and analytical conditions can all be printed out. | |
Instrument self-check | Check current status of each function key. | |
Characteristic Concentration and Detection Limit | Air-C2H2 flame | Cu: Characteristic concentration≤0.025mg/L, Detection limit≤0.006mg/L; |
Function Expansion | Hydride vapor generator can be connected for hydride analysis. | |
Dimensions and Weights | 1020(L)´490(W)´540(H)mm, unpacked 80Kg | |