| Place of Origin: | Jiangsu China (Mainland) |
| Model Number: | CS-8820 |
Analyzer is widely used in various metals material, metallurgy,concrete, mining, glasses and porcelain, petrol oil etc.
1. Summary
CS-8820 Type High-frequency Infrared Carbon & Sulfur Analyzer compatible
applying with WF-T88 Type High frequency inductive combustion furnace, can
exactly measure the carbon and sulfur mass fraction in steel, ion, alloy,
casting core sand, nonferrous metals and other materials very quickly. This set
of equipment is a high-tech products integrate light, motor, electronic,
computer and analysis technology, with features of wide measuring scope, strong
anti interference, multiple function, easy to operate, accurate and reliable
analysis result.
2. Technical specification and index
2.1. Basic parameter Equipment performance and accessories
Analyzing ElementCarbon and sulfur measure
Analyzing principleHigh frequency inductive combustion furnacedetection by
infrared absorption.
Measure Range:Carbon0.001%-6.0000% (can extend to 99.999%);
Sulfur0.001%-0.5000% (can extend to 99.999%)
Sample weighting amount
Solid steel scaleStandard 0.5g
Analysis Error:Carbonaccord with ISO9556 standard;
Sulfuraccord with ISO9556 standard
SensitivityMinimal amountC/S0.1ppm
Analyzing time20-100 seconds adjustableusually 35 seconds
Working period:24 hours continuously operation
High frequency inductive combustion furnace1. High frequency inductive
combustion furnace output power:2200WFrequency20MHz.
2. Auto-checking: Automatic overtime/over flow alarming system.
3.The optional current/voltage/power regulator for furnace temperature
control; it is applicable to samples of various materials.
4. Gas paththe high precision flow controller ensures the stability of gas
flow as well the gas intake system.
Dust filter:0.4um submicron metal filter secures thorough separation of
dust from gas.
Furnace head heating Increase conversion rate of sulfur, to get a steady
analysis result.
De-dusting system:The combustion head self-cleaning devices effective for
reducing the influence of dust for the result of analysis; ash removing
system for the inlet.
Detection system:1. Circuit design: the modularized design with double CPU
for the whole frame is divided into the upper and lower levels, the lower
one adopt Atmega162 to control the single machine, while the electronic
circuits are highly integrated stably;
2.SamplingAdopt high precision 24 leave ADS124 sampling CMOS chips,;
3.Communicationthe upper and lower machine adopt USB terminal to increase
the communicating speed;
4.ConnectionInfrared detection parts and High frequency inductive
combustion furnace connect with fiber, assorted with series shelter
isolated circuit, avoid high frequency interfere;
5. Light sourcethe tailor-made new model platinum infrared light source
has lasting caloricity and efficient spectral property
6.Analysis cell: the gilded carbon & sulfur analysis cell is fit with the
high precision TSC infrared detector (1011);
7. Motor: the special-purpose aviation modulation motor is applied with
satisfactory thermal stability whose life expectancy exceeds 100000 hours.
Instrument diameter and weight:Burning partsW70×H80×D55cm Weight60kg
Inspecting partsW55×H28×D45cm Weight18kg
Authorized standardISO9556ISO4935
Operation system configuration Detailed information refers to the scope of
supply.
2.2. Analyzing software
Operating software:Adopt Deliph software,WINDOWS XP full Chinese operation
interface.
Analyzing route way:With channel management function, carbon and sulfur
channel can increase, delete and edit separately without quantity limit.
Preset low carbon and high sulfur interface.
Analyzing function:Dynamic data accumulation when analyzing, 20times
sampling per second, increasing the sensitivity and accuracy.
With sample management function, can edit the name, mark of sample, also
can do adding and deleting sample name.
Display function:Separately curve frame for Carbon & Sulfur, dynamic show
the data and Carbon & Sulfur release curve.
Data processing function:Analysis results store by ACCESS data-base way,
it can store data and curve every time; can search the results according
to time, operator, sample name and mark any time; with curve/data storage,
deduction of blankness, preset of parameters, channel selection, data
statistic, result adjustment and curve preparing and other function. Can
realize creating carbon and sulfur working curve and do the curve fitting
in the software.
Printing function:Printing mode diversification: Laboratory and inspection
station printing mode and can self-design proper mode.
Self-diagnose function:System self-diagnose function,can realize checking
furnace head and sealing performance of the gas chamber by the soft ware.
3. Scope of Supply
One set of CS-8820 Carbon & Sulfur Analyzer includes the following components:
ItemComponent partsQuantity
1Main body of CS-8820 Type High-frequency Infrared Carbon & Sulfur
Analyzer1
2WF-T88 Type High frequency inductive combustion furnace1
3SDSL (Sartorius) BS124S Electronic balance1
4Tsinghua Tongfang commercial micro computer1
5P1007 HP laser printer1
| Business Type: | Manufacturer, Trading Company |
| Product/Service (We Sell): | caplamp,mining light,cordless cap lamp,analyzer,led light,led mining,led light,searchlight,analysis instrument,Spectrophotometer |
| Brands: | JIEBO |
| Number of Employees: | 51 - 100 People |
| Main Markets: | North America South America Eastern Europe Southeast Asia Africa Oceania Mid East Eastern Asia Western Europe |
| Total Annual Sales Volume: | US$5 Million - US$10 Million |
| Export Percentage: | 31% - 40% |
| Factory Size (Sq.meters): | 10,000-30,000 square meters |
| Factory Location: | Qianqiao Town Industrial Concentration Area (Xinan Village) |
| QA/QC: | In House |
| Number of Production Lines: | 5 |
| Number of R&D Staff: | 5 - 10 People |
| Number of QC Staff: | 5 - 10 People |
| Management Certification: | ISO 9000/9001/9004/19011: 2000 |
| Contract Manufacturing: | OEM Service Offered |