1,Pressing depth:30 mm~65 mm adjustable,one type:5 mm,error:±2 mm; |
2,Pressing release rate:1:1 ; |
3,pressure relief valve:0.65Mpa;error:±0.12Mpa; |
4,Pressure of gas road is no less than 1Mpa; |
5, Ventilation quanity:500ML no adjust,error:±10%; |
working pressure:≤816.4N ; |
6,input power(DC):≤17W; AC:≤54W; |
7,Press the ventilation ratio:30:2 continuously adjustable;,error:±1 time; |
8,Pressing frequency: 100times/ min,110times/ min,120times/ min error:±1 time; |
9,The fastness of fastening band fastness is not less than 100N; |
10,The thickness range of patients was 155 to 305MM. |
11,The pressure is not enough to alarm, the pressure depth is reduced automatically, and the manual recovery after air pressure recovery must be deep. |
12,Electricity, according to the low voltage alarm, start the suspended function; |
13,Default mode:Press the ventilation ratio: 30: 2,press depth 30mm,press frequence 100 pcs/min Ventilation quanity 500ML |
14,Safety requirements refer to medical electrical appliances safety standard GB 9706.1-2007, electromagnetic compatibility standard YY05052012; |
15,Earthquake Resistance |
16,drop resistance |
17,The enclosure protection level is IP55 (dustproof and water spray); |
18,noise≤70dB ; |
19,Working environment : a)Temperature:-20℃~55℃; |
b)Relative temperatur:10%~90%; |
c)Atmospheric pressure:500hPa~1060hPa. |